Density Meters / Specific Gravity Analyzer

Specific Gravity Analyzer DC-54
DC-54

- Direct gas measurement in pipeline
- Automatic temperature compensation
- Increased safety level
- ATEX Hazardous area certification
- Easy installation, low maintenance

Overview
DC-54 Specific Gravity Analyzer is another instrument of LEMIS process DC series instruments, designed for accurate, continuous monitoring of the technological process. High-accuracy measured data for natural gas mixtures are required in entire range of pressure, temperature and compositions encountered in custody transfer. The gas is passed over the cylinder tube resonance sensor vibrating at its natural frequency. The mass of gas in contact with the vibrating walls vibrates with the cylinder. Depends upon the gas density and, since increasing the vibrating mass decreases the natural frequency of vibration, the gas density is simply determined by measuring the change in sensor resonant frequency. This measurement method provides precise gas specific gravity values.
Documentation

Please log in or register with LEMIS Process to get access to following documents:
file BROCHURE Specific Gravity Meter DC-53 (DC_54_SG_curv.pdf, 2.57 MB)
Specifications

 

Specific Gravity

0 to 2

Specific Gravityy Accuracy

±0.1% of reading

Specific Gravity Repeatability

±0.02% of reading

Specific Gravity Resolution

0.0001

Pressure Range

Up to 30 Bar

Temperature Range

-40°C to +85°C (-40°F to +185°F)

Temperature Accuracy

±0.5°C (±0.9°F)

Ambient temperature

-20°C to +60°C (-4°F to +140°F)

Process Connection

NPT Thread 1/8” or 1/4"

Applications
- Marine
- Military applications
- Pharmaceutical, cosmetic industries
- Petroleum industry
- Food & Beverages